Implement patient vital signs monitoring system with Contec CMS7000PLUS integration and real-time dashboard
This commit is contained in:
@@ -0,0 +1,381 @@
|
||||
import logging
|
||||
from datetime import datetime, timezone
|
||||
from typing import Optional, List
|
||||
import struct
|
||||
|
||||
from schemas import NormalizedVitals
|
||||
|
||||
logger = logging.getLogger(__name__)
|
||||
|
||||
# Contec CMS7000PLUS / CMS8000 / CMS9200PLUS OBX identifiers mapping to schema fields
|
||||
CONTEC_PARAM_MAP = {
|
||||
# Heart Rate / Pulse Rate
|
||||
"HR": "heart_rate",
|
||||
"HEART_RATE": "heart_rate",
|
||||
"ECG_HR": "heart_rate",
|
||||
"PR": "heart_rate",
|
||||
"PULSE_RATE": "heart_rate",
|
||||
"SPO2_PR": "heart_rate",
|
||||
"ECGHR": "heart_rate",
|
||||
"ECG-HR": "heart_rate",
|
||||
"PULSERATE": "heart_rate",
|
||||
|
||||
# SpO2
|
||||
"SPO2": "spo2",
|
||||
"SAO2": "spo2",
|
||||
"SPO2_PCT": "spo2",
|
||||
"OXYGEN_SAT": "spo2",
|
||||
"SPO2_VAL": "spo2",
|
||||
"O2SAT": "spo2",
|
||||
|
||||
# Blood Pressure (NIBP)
|
||||
"NIBP_SYS": "systolic_bp",
|
||||
"NIBP_DIA": "diastolic_bp",
|
||||
"NIBP_MEAN": "map_bp",
|
||||
"NIBP_MAP": "map_bp",
|
||||
"SYS": "systolic_bp",
|
||||
"DIA": "diastolic_bp",
|
||||
"MAP": "map_bp",
|
||||
"NBP_S": "systolic_bp",
|
||||
"NBP_D": "diastolic_bp",
|
||||
"NBP_M": "map_bp",
|
||||
"NIBP-S": "systolic_bp",
|
||||
"NIBP-D": "diastolic_bp",
|
||||
"NIBP-M": "map_bp",
|
||||
"BP_SYS": "systolic_bp",
|
||||
"BP_DIA": "diastolic_bp",
|
||||
"BP_MEAN": "map_bp",
|
||||
|
||||
# Respiratory Rate
|
||||
"RESP": "respiratory_rate",
|
||||
"RESP_RATE": "respiratory_rate",
|
||||
"RR": "respiratory_rate",
|
||||
"BR": "respiratory_rate",
|
||||
"RESPRATE": "respiratory_rate",
|
||||
"RESP-RATE": "respiratory_rate",
|
||||
|
||||
# Temperature
|
||||
"TEMP": "temperature",
|
||||
"TEMP1": "temperature",
|
||||
"TEMP2": "temperature",
|
||||
"T1": "temperature",
|
||||
"T2": "temperature",
|
||||
"BODY_TEMP": "temperature",
|
||||
"TEMP-1": "temperature",
|
||||
"TEMP-2": "temperature",
|
||||
"BODYTEMP": "temperature",
|
||||
}
|
||||
|
||||
# Sentinel value used by Contec to mark an invalid/disconnected parameter
|
||||
INVALID_SENTINEL_U16 = 9999 # 0x270F
|
||||
INVALID_SENTINEL_F32 = 9999.0
|
||||
|
||||
|
||||
def safe_float(value: str) -> Optional[float]:
|
||||
try:
|
||||
return float(value)
|
||||
except (ValueError, TypeError):
|
||||
return None
|
||||
|
||||
|
||||
def is_valid_u16(val: Optional[int]) -> bool:
|
||||
return val is not None and val != INVALID_SENTINEL_U16
|
||||
|
||||
|
||||
def is_valid_f32(val: Optional[float]) -> bool:
|
||||
return val is not None and abs(val - INVALID_SENTINEL_F32) > 1.0
|
||||
|
||||
|
||||
def read_u16_le(buf: bytes, offset: int) -> Optional[int]:
|
||||
if offset + 2 <= len(buf):
|
||||
return struct.unpack_from('<H', buf, offset)[0]
|
||||
return None
|
||||
|
||||
|
||||
def read_f32_le(buf: bytes, offset: int) -> Optional[float]:
|
||||
if offset + 4 <= len(buf):
|
||||
return struct.unpack_from('<f', buf, offset)[0]
|
||||
return None
|
||||
|
||||
|
||||
def parse_contec_hl7_text(raw_text: str, client_ip: str) -> Optional[NormalizedVitals]:
|
||||
"""
|
||||
Lightweight, native parser for Contec HL7 messages.
|
||||
Splits segments and fields manually, avoiding third-party library dependencies.
|
||||
"""
|
||||
try:
|
||||
# Strip MLLP framing if present (VT=0x0b, FS=0x1c, CR=0x0d)
|
||||
clean_text = raw_text.strip()
|
||||
if clean_text.startswith("\x0b"):
|
||||
clean_text = clean_text[1:]
|
||||
if "\x1c\x0d" in clean_text:
|
||||
clean_text = clean_text.split("\x1c\x0d")[0]
|
||||
|
||||
# Split into segments by CR or LF
|
||||
segments = [seg.strip() for seg in clean_text.replace("\n", "\r").split("\r") if seg.strip()]
|
||||
|
||||
if not segments or not segments[0].startswith("MSH"):
|
||||
return None
|
||||
|
||||
# Parse MSH segment
|
||||
msh_fields = segments[0].split("|")
|
||||
device_id = msh_fields[2] if len(msh_fields) > 2 and msh_fields[2] else f"CMS7000_{client_ip}"
|
||||
|
||||
patient_id = "UNKNOWN"
|
||||
vitals_dict = {
|
||||
"device_id": device_id,
|
||||
"patient_id": patient_id,
|
||||
"timestamp": datetime.now(timezone.utc),
|
||||
}
|
||||
|
||||
found_any = False
|
||||
|
||||
# Parse PID and OBX segments
|
||||
for seg in segments[1:]:
|
||||
fields = seg.split("|")
|
||||
if not fields:
|
||||
continue
|
||||
|
||||
seg_name = fields[0]
|
||||
|
||||
if seg_name == "PID":
|
||||
if len(fields) > 3 and fields[3]:
|
||||
patient_id = fields[3].split("^")[0]
|
||||
vitals_dict["patient_id"] = patient_id
|
||||
|
||||
elif seg_name == "OBX":
|
||||
if len(fields) > 5:
|
||||
obs_id = fields[3].split("^")[0].strip().upper()
|
||||
obs_val = fields[5].strip()
|
||||
|
||||
if not obs_val or obs_val == "---":
|
||||
continue
|
||||
|
||||
val_float = safe_float(obs_val)
|
||||
if val_float is None:
|
||||
continue
|
||||
|
||||
for param_key, mapped_field in CONTEC_PARAM_MAP.items():
|
||||
if obs_id == param_key or param_key in obs_id:
|
||||
vitals_dict[mapped_field] = val_float
|
||||
found_any = True
|
||||
logger.info(f"Parsed Contec HL7 Vital: {mapped_field} = {val_float} (from {obs_id})")
|
||||
break
|
||||
|
||||
return NormalizedVitals(**vitals_dict) if found_any else None
|
||||
|
||||
except Exception as e:
|
||||
logger.error(f"Error parsing Contec HL7 text: {e}")
|
||||
return None
|
||||
|
||||
|
||||
def _extract_vitals_block(data: bytes, offset: int, client_ip: str) -> Optional[NormalizedVitals]:
|
||||
"""
|
||||
Extract vitals from a Contec vitals block starting at the given offset.
|
||||
|
||||
Known Contec CMS7000PLUS binary vitals block layout (confirmed from packet captures):
|
||||
|
||||
PACKET TYPE B (header bytes [xx][00][04][46] or [xx][yy][04][46]):
|
||||
-----------------------------------------------------------------------
|
||||
Bytes 8-11 : float32 LE → NIBP Systolic (9999.0 = invalid)
|
||||
Bytes 12-15 : float32 LE → NIBP Diastolic (9999.0 = invalid)
|
||||
Bytes 16-19 : float32 LE → NIBP MAP (9999.0 = invalid)
|
||||
Bytes 20-23 : float32 LE → Temperature 1 (°C, 9999.0 = invalid)
|
||||
Bytes 24-27 : float32 LE → Temperature 2 (°C, 9999.0 = invalid)
|
||||
|
||||
Vitals summary block (at data[offset] position, found by scanning for SpO2+HR pattern):
|
||||
Offset +0 : uint16 LE → SpO2 (%) (9999 = invalid)
|
||||
Offset +2 : uint16 LE → Heart Rate (bpm)(9999 = invalid, ECG not connected)
|
||||
Offset +4 : uint16 LE → Pulse Rate (bpm)(from SpO2 probe, also 9999 if invalid)
|
||||
Offset +6 : uint16 LE → (SpO2 alarm/threshold)
|
||||
Offset +8 : uint16 LE → (mode flags)
|
||||
Offset +10 : uint16 LE → NIBP Systolic (9999 = invalid)
|
||||
Offset +12 : uint16 LE → NIBP Diastolic (9999 = invalid)
|
||||
"""
|
||||
vitals_dict = {
|
||||
"device_id": f"CMS7000PLUS_{client_ip}",
|
||||
"patient_id": "UNKNOWN",
|
||||
"timestamp": datetime.now(timezone.utc),
|
||||
}
|
||||
found = False
|
||||
|
||||
spo2 = read_u16_le(data, offset)
|
||||
hr_ecg = read_u16_le(data, offset + 2)
|
||||
hr_spo2 = read_u16_le(data, offset + 4)
|
||||
|
||||
# SpO2 value: valid range 50–100
|
||||
if spo2 is not None and 50 <= spo2 <= 100:
|
||||
vitals_dict["spo2"] = float(spo2)
|
||||
found = True
|
||||
|
||||
# Heart rate: prefer ECG HR; fall back to SpO2-derived PR if ECG invalid
|
||||
if is_valid_u16(hr_ecg) and 20 <= hr_ecg <= 300:
|
||||
vitals_dict["heart_rate"] = float(hr_ecg)
|
||||
found = True
|
||||
elif is_valid_u16(hr_spo2) and 20 <= hr_spo2 <= 300:
|
||||
vitals_dict["heart_rate"] = float(hr_spo2)
|
||||
found = True
|
||||
|
||||
return NormalizedVitals(**vitals_dict) if found else None
|
||||
|
||||
|
||||
def _parse_float_vitals(data: bytes, base_offset: int, vitals_dict: dict) -> bool:
|
||||
"""
|
||||
Parse float32 vitals (NIBP, Temperature) from packet starting at base_offset.
|
||||
Returns True if any valid vitals found.
|
||||
"""
|
||||
found = False
|
||||
|
||||
nibp_sys = read_f32_le(data, base_offset)
|
||||
nibp_dia = read_f32_le(data, base_offset + 4)
|
||||
nibp_map = read_f32_le(data, base_offset + 8)
|
||||
temp1 = read_f32_le(data, base_offset + 12)
|
||||
temp2 = read_f32_le(data, base_offset + 16)
|
||||
|
||||
if nibp_sys is not None and is_valid_f32(nibp_sys) and 50.0 < nibp_sys < 300.0:
|
||||
vitals_dict["systolic_bp"] = round(nibp_sys, 1)
|
||||
found = True
|
||||
|
||||
if nibp_dia is not None and is_valid_f32(nibp_dia) and 20.0 < nibp_dia < 200.0:
|
||||
vitals_dict["diastolic_bp"] = round(nibp_dia, 1)
|
||||
found = True
|
||||
|
||||
if nibp_map is not None and is_valid_f32(nibp_map) and 20.0 < nibp_map < 250.0:
|
||||
vitals_dict["map_bp"] = round(nibp_map, 1)
|
||||
found = True
|
||||
|
||||
if temp1 is not None and is_valid_f32(temp1) and 30.0 < temp1 < 45.0:
|
||||
vitals_dict["temperature"] = round(temp1, 1)
|
||||
found = True
|
||||
elif temp2 is not None and is_valid_f32(temp2) and 30.0 < temp2 < 45.0:
|
||||
vitals_dict["temperature"] = round(temp2, 1)
|
||||
found = True
|
||||
|
||||
return found
|
||||
|
||||
|
||||
def _find_vitals_block_in_stream(data: bytes, client_ip: str) -> Optional[NormalizedVitals]:
|
||||
"""
|
||||
Scan the data stream for Contec CMS7000PLUS sub-packets.
|
||||
|
||||
Contec packets follow this framing pattern:
|
||||
[len_lo] [len_hi] [04] [46] [sub_type] [00] [sub_len_lo] [00] ...payload...
|
||||
|
||||
OR the data is a concatenation of multiple sub-packets separated by
|
||||
[xx][xx][04][47] (end marker?)
|
||||
|
||||
The key is to find [04][46] marker bytes and parse the packet.
|
||||
"""
|
||||
vitals_dict = {
|
||||
"device_id": f"CMS7000PLUS_{client_ip}",
|
||||
"patient_id": "UNKNOWN",
|
||||
"timestamp": datetime.now(timezone.utc),
|
||||
}
|
||||
found_any = False
|
||||
|
||||
# Scan through the buffer looking for [04 46] packet markers
|
||||
i = 0
|
||||
while i < len(data) - 8:
|
||||
# Look for the [04][46] marker which is the packet type indicator
|
||||
if data[i + 2] == 0x04 and data[i + 3] == 0x46:
|
||||
pkt_len = struct.unpack_from('<H', data, i)[0] # little-endian length
|
||||
|
||||
# Packet type B: small packets (< 600 bytes) with vitals summary
|
||||
# These have float32 NIBP values at offset +8 relative to packet start
|
||||
if 30 < pkt_len < 600 and i + 8 <= len(data):
|
||||
# Read float32 vitals at offsets 8, 12, 16, 20, 24 from packet start
|
||||
_parse_float_vitals(data, i + 8, vitals_dict)
|
||||
|
||||
# Find the vitals summary block in this packet
|
||||
# It appears near the tail of the resp waveform data
|
||||
# The pattern we look for: SpO2 byte (50-100), followed by various u16 values
|
||||
# In packet B, the summary is 75 bytes before the end
|
||||
if pkt_len > 50:
|
||||
# The summary block is consistently found around offset 322 in 397-byte packets
|
||||
# Relative to packet start: around pkt_len - 75
|
||||
summary_offset = i + max(4, pkt_len - 80)
|
||||
if summary_offset + 10 < len(data):
|
||||
# Scan in the last ~100 bytes of the packet for SpO2 pattern
|
||||
scan_end = min(i + pkt_len, len(data) - 4)
|
||||
scan_start = max(i + 4, scan_end - 100)
|
||||
for j in range(scan_start, scan_end - 4, 2):
|
||||
spo2 = read_u16_le(data, j)
|
||||
hr = read_u16_le(data, j + 2)
|
||||
pr = read_u16_le(data, j + 4)
|
||||
|
||||
# Check if this looks like [SpO2][HR/sentinel][PR]
|
||||
spo2_ok = spo2 is not None and 50 <= spo2 <= 100
|
||||
# HR can be 9999 (invalid/ECG disconnected) or valid 20-300
|
||||
hr_or_sentinel = hr is not None and (hr == INVALID_SENTINEL_U16 or 20 <= hr <= 300)
|
||||
pr_ok = pr is not None and (pr == INVALID_SENTINEL_U16 or 20 <= pr <= 300)
|
||||
|
||||
if spo2_ok and hr_or_sentinel and pr_ok:
|
||||
vitals_dict["spo2"] = float(spo2)
|
||||
found_any = True
|
||||
|
||||
# Use ECG HR if valid, else use SpO2 PR
|
||||
if hr is not None and hr != INVALID_SENTINEL_U16 and 20 <= hr <= 300:
|
||||
vitals_dict["heart_rate"] = float(hr)
|
||||
elif pr is not None and pr != INVALID_SENTINEL_U16 and 20 <= pr <= 300:
|
||||
vitals_dict["heart_rate"] = float(pr)
|
||||
|
||||
logger.info(
|
||||
f"Parsed vitals block at offset {j}: "
|
||||
f"SpO2={spo2}, HR={hr}, PR={pr}"
|
||||
)
|
||||
break
|
||||
|
||||
# Move past this packet
|
||||
i += max(pkt_len, 4)
|
||||
else:
|
||||
i += 1
|
||||
|
||||
return NormalizedVitals(**vitals_dict) if found_any else None
|
||||
|
||||
|
||||
def parse_contec_binary_packet(data: bytes, client_ip: str) -> Optional[NormalizedVitals]:
|
||||
"""
|
||||
Decodes the Contec CMS7000PLUS proprietary binary protocol.
|
||||
|
||||
The monitor sends two alternating TCP packets:
|
||||
- Packet A (991 bytes): ECG waveform + SpO2 waveform + vitals footer
|
||||
- Packet B (397 bytes): Resp waveform + vitals summary + param block
|
||||
|
||||
Both have the structure: [len_lo][len_hi][04][46][subtype][00][sublen][00]...[payload]...
|
||||
"""
|
||||
try:
|
||||
return _find_vitals_block_in_stream(data, client_ip)
|
||||
except Exception as e:
|
||||
logger.error(f"Error parsing Contec binary packet: {e}")
|
||||
return None
|
||||
|
||||
|
||||
def parse_contec_data(raw_data: bytes, client_ip: str) -> Optional[NormalizedVitals]:
|
||||
"""
|
||||
Unified entry point for parsing data from a Contec CMS7000PLUS monitor.
|
||||
Attempts HL7 text parsing first, falling back to proprietary binary parsing.
|
||||
"""
|
||||
if not raw_data:
|
||||
return None
|
||||
|
||||
# Attempt 1: HL7 Text Decoding
|
||||
try:
|
||||
text = raw_data.decode("utf-8", errors="ignore").strip()
|
||||
if "MSH" in text or "\x0bMSH" in text:
|
||||
vitals = parse_contec_hl7_text(text, client_ip)
|
||||
if vitals:
|
||||
return vitals
|
||||
except Exception:
|
||||
pass
|
||||
|
||||
# Attempt 2: Binary Parsing (Contec proprietary protocol)
|
||||
vitals = parse_contec_binary_packet(raw_data, client_ip)
|
||||
if vitals:
|
||||
return vitals
|
||||
|
||||
# Logging fallback
|
||||
logger.warning(
|
||||
f"Data packet from {client_ip} unrecognized. "
|
||||
f"Len: {len(raw_data)}, Hex head: {raw_data[:32].hex()}"
|
||||
)
|
||||
return None
|
||||
Reference in New Issue